Gamma dose rate and shielding
Point-source dose rate from the emission spectrum, with attenuation and buildup, and the shield thickness solved backwards.
No sign-in needed. Nothing you type leaves your browser.
How the dose rate is obtained
Most calculators of this kind look up a gamma constant in a table. This one computes it. For each photon line of the nuclide, the energy and the emission probability come from the evaluated nuclear data, and the mass energy-absorption coefficient of air at that energy comes from the NIST tables. Summing E·y·(µen/ρ) over the lines and dividing by 4π gives the air kerma rate constant directly. The result agrees with published constants to within about 2% for the well-characterised nuclides.
Computing rather than copying has a practical benefit beyond provenance: the low-energy cutoff becomes a control you can move. Published constants apply a cutoff, usually 20 keV, to exclude photons that the source encapsulation absorbs before they reach anyone. That choice is invisible in a table and decisive in the answer. For Am-241 a 10 keV cutoff instead of 20 keV raises the constant by roughly a factor of eight, because the 17 keV neptunium L X-rays are absorbed strongly in air and dominate the kerma integral while contributing nothing at a metre through any real capsule.
Attenuation, and why it is not enough
A shield removes photons following exp(−µx), where µ is the linear attenuation coefficient — the mass attenuation coefficient at that energy multiplied by the density. The calculator applies this line by line, so a nuclide with a broad spectrum hardens correctly as it passes through: the low-energy lines are removed preferentially and the average energy behind the shield is higher than in front of it.
Exponential attenuation alone describes a narrow beam, where any photon that interacts is gone. A real shield is a broad beam. Photons Compton-scatter inside the material, change direction, and a substantial fraction still emerge toward the receptor. Ignoring them can underestimate the dose rate behind a thick shield by a factor of several. The half-value layers this tool reports are therefore narrow-beam values, noticeably thinner than the broad-beam figures in shielding handbooks — 5.5 mm rather than about 6.5 mm of lead at 662 keV.
Buildup factors
The correction is the buildup factor B, the ratio of the true dose rate to the attenuated-only dose rate. It grows with the number of mean free paths and depends on material, energy and geometry. This calculator implements the two-parameter Berger form B = 1 + a·µx·eb·µx but ships no coefficients. The most widely used tabulation, the geometric-progression fit, is the body of a paid consensus standard, and reproducing it here would not be legitimate. Enter a and b from the reference you are working to, and the tool will show what difference they make.
With attenuation only, the shield thickness this calculator returns is a lower bound. Treat it as the starting point for a design, never as the design.
Geometry assumptions
A point source in air, radiating isotropically, with a receptor at a stated distance and the shield between them. No room scatter, no skyshine, no source self-absorption, no capsule, and no correction from air kerma to a dose quantity in tissue. Real sources are extended, and close to an extended source the inverse-square falloff does not hold — near a large plane the dose rate barely falls with distance at all.