Europium-152
Eu-152 · Europium, Z = 63, A = 152
Europium-152 decays by electron capture and beta-minus emission, half-life 13.517 years. Specific activity is 6.44e+12 Bq/g (174 Ci/g), so a gigabecquerel comes to 155 µg — weighable, but on an analytical balance. Over a year the activity falls to 95.00%, and over forty years to 12.86%.
An air kerma rate constant of 0.151 mGy·m²/(GBq·h) is 2.0× Cs-137 and 2.0× less than Co-60, and 12 of 96 among the photon emitters carried here.
The two columns of the line table disagree here: the 1408.0 keV line carries the largest share of the dose rate although the 344.3 keV line is emitted more often — energy, not frequency, decides which photons set the shield.
This is a shield that has to be designed: 6.37 mm of lead for a factor of two and 27.8 mm for a factor of ten, or 12.4 mm of steel to halve it, at which point the mass of the shield is part of the problem. Reaching 20 µSv/h from 1 GBq at a metre takes 23.9 mm of lead. The tenth-value layer runs 4.4 times the half-value layer, not the 3.32 a single energy would give.
The beta endpoint is 1.47 MeV, mean 0.298 MeV over 12 branches, which between them occur in 27.9% of decays; that endpoint comes from a branch of only 8.24%, while 13.73% of decays take the 0.696 MeV branch. Shielding is sized on the endpoint, dose on the mean. That endpoint stops in 5.58 mm of acrylic or 2.66 mm of glass. Of the beta energy, 0.34% turns into X-rays in acrylic and 4.23% in lead.
Eu-152 for detector efficiency calibration and in activated concrete
Eu-152 emits strong photon lines from tens of kiloelectronvolts to beyond a megaelectronvolt, so a single source covers much of the energy range over which a gamma spectrometer's efficiency curve is built. Its half-life, listed in the first table below, keeps a calibration source serviceable for a decade or more. It is made by neutron capture in Eu-151, and capture in Eu-153, the other natural isotope, gives Eu-154 alongside it.
Most decays are electron captures to samarium, but a substantial minority are beta-minus decays to gadolinium, which give several prominent gamma lines in the photon table, while the X-rays listed there are samarium X-rays that follow electron capture. The same richness of lines causes errors at close geometry: several lines are emitted in cascade, and coincidence summing removes counts from one peak and adds them to a sum peak. An efficiency curve taken with the source on the detector cap will be wrong unless summing corrections are applied or the source is counted at a distance. On a detector sensitive to low-energy photons the X-rays from electron capture sum with gamma rays and with each other, so correction codes must handle every such coincidence.
Neutron capture in the trace europium of concrete also produces Eu-152, so with Eu-154 it sets part of the dose rate when a reactor bioshield is dismantled. The shielding table here gives layers for the whole spectrum, dominated by the harder lines. The europium is only a trace, but its capture cross-section for thermal neutrons is very large, and reactor graphite is activated in the same way. Drilled cores measured along their length give the activation depth, and with it the volume of concrete to be managed as radioactive waste. Co-60 decays faster than either europium nuclide, so the longer dismantling is deferred, the larger the europium share of the gamma activity left in the concrete.
Uranium- and thorium-series nuclides in the concrete itself emit lines that a germanium detector cannot resolve from some of the strongest Eu-152 lines, and Eu-154 has an intense line just above the main low-energy Eu-152 gamma. At low activity concentrations a figure taken from the high-energy peaks that coincide with natural lines overstates Eu-152 unless the natural contribution is estimated from other lines of the same series and subtracted. The low-energy peak needs a fit that separates the Eu-154 line.
- IAEA Live Chart of Nuclides (ENSDF decay data)
- LNHB / DDEP, Recommended data: Eu-152 decay tables
- IAEA Technical Reports Series No. 389, Radiological Characterization of Shut Down Nuclear Reactors for Decommissioning Purposes
- Jonsson, Kastlander, Vidmar and Ramebäck, Experimental validation of correction factors for γ–γ and γ–X coincidence summing of Ba-133, Eu-152 and Sb-125 in volume sources, Journal of Radioanalytical and Nuclear Chemistry (2020)
Half-life, specific activity, dose rate and beta energies
| Half-life | 13.517 years (4.266e+8 s) |
| Decay mode | electron capture and beta-minus decay |
| Specific activity | 6.44e+12 Bq/g (174 Ci/g) |
| Air kerma rate constant Γ (δ = 20 keV) | 0.151 mGy·m²/(GBq·h) |
| Dose rate, 1 GBq at 1 m | 0.151 mGy/h |
| Dose rate, 1 Ci at 1 m | 5.58 mGy/h |
| Kerma-weighted mean photon energy | 934.1 keV |
| Beta endpoint / mean | 1.47 MeV / 0.298 MeV |
56 lines above the cutoff, and what each contributes
2 further lines below the 20 keV cutoff, the highest at 6.86 keV and 14.2% emission probability in all, are excluded here and from Γ. Probability and dose share.
| Energy (keV) | Emission probability (%) | Share of dose rate (%) |
|---|---|---|
| 1408.01 | 20.87 | 23.16 |
| 1112.08 | 13.67 | 12.64 |
| 964.06 | 14.51 | 11.95 |
| 1085.84 | 10.11 | 9.17 |
| 778.90 | 12.93 | 8.86 |
| 344.28 | 26.59 | 8.11 |
| 867.38 | 4.23 | 3.18 |
| 40.12 | 37.8 | 3.13 |
| 121.78 | 28.53 | 2.55 |
| 39.52 | 20.87 | 1.78 |
| 1299.14 | 1.633 | 1.71 |
| 1089.74 | 1.734 | 1.58 |
| 244.70 | 7.55 | 1.56 |
| 1212.95 | 1.415 | 1.40 |
Showing the 14 largest contributors of 56 lines above the cutoff; the remainder together carry 9.24% of the dose rate.
6.37 mm of lead halves this spectrum
Solved numerically across all 56 lines, narrow beam. Why not one representative energy.
| Material | HVL (mm) | TVL (mm) | TVL / HVL |
|---|---|---|---|
| lead | 6.37 | 27.8 | 4.37 |
| tungsten | 4.07 | 17.4 | 4.28 |
| iron | 12.4 | 46.9 | 3.79 |
| copper | 11.0 | 41.8 | 3.82 |
| concrete | 40.1 | 149 | 3.71 |
| water | 86.5 | 314 | 3.64 |
| aluminum | 36.0 | 134 | 3.72 |
A single energy would give 3.32. Spectrum and attenuation ratios.
5.58 mm of acrylic stops the 1.47 MeV endpoint
Katz–Penfold fit to the 1.47 MeV endpoint — a stopping thickness rather than an attenuation length. Why the material matters more than the thickness.
| Absorber | Range for the endpoint (mm) |
|---|---|
| acrylic | 5.58 |
| aluminum | 2.46 |
| water | 6.64 |
| glass | 2.66 |
| Shield material | Fraction of beta energy converted to X-rays |
|---|---|
| acrylic (Z ≈ 6.5) | 0.34% |
| lead (Z = 82) | 4.23% |
Activity over decades and centuries
Ten half-lives is 135 years, so forty years of decay stops short of the 0.1% mark: it removes most of the activity but leaves 12.86%. The mean life 1/λ, the quantity that enters an integrated dose, is 19.5 years.
| Elapsed | Fraction remaining |
|---|---|
| 1 half-life | 50.0 % |
| 2 half-lives | 25.0 % |
| 5 half-lives | 3.13 % |
| 10 half-lives | 0.0977 % |
| Time to fall to 10 % of today's activity | 44.9 years |
| Time to fall to 1 % | 89.8 years |
| Time to fall to 0.1 % | 135 years |
Limits of these dose rates
- 5.58 mm of acrylic for the 1.47 MeV endpoint: a fitted depth of penetration, not a skin dose.
- Air kerma, 1 µGy counted as 1 µSv — not H*(10).
- Model limits and internal dose.
Gamma, beta and decay calculators for Eu-152
- Gamma dose rate and shielding
- Beta dose rate and shielding
- Decay and half-life
- Mass and activity
- Detection limits (MDA / MDC)
Other Europium nuclides: Eu-152m1, Eu-152m2, Eu-154, Eu-154m
Computed from the IAEA Nuclear Data Section — Livechart API (ENSDF) and the NIST X-Ray Mass Attenuation Coefficients. Derivations and citations.