Minimum detectable activity: worked example

Minimum detectable activity: an MDA calculation worked step by step

By Radiation Lab · Published · Reviewed

You counted a sample and no peak stood out. What is the smallest activity that count could reliably have seen? That figure is the minimum detectable activity, and this note works it through for Cs-137 on a gamma spectrometer, line by line, using the equations and the assumptions of the MDA calculator: Currie's detection limits with a paired blank. Then it reads three possible results against the limits, since that is where most confusion starts.

The measurement

The inputs below are illustrative but typical of a shielded detector and a one-hour count.

InputValue
Blank count rate in the 662 keV peak region2 cpm
Count time t, same for sample and blank3600 s
Full-energy peak efficiency ε3.0%
Emission probability Y, IAEA Livechart85.1%
Sample mass0.5 kg
Coverage k, for 5% false positives and 5% false negatives1.645

Step 1: the critical level LC

The expected blank count is B = 2 × 60 = 120 counts. A net count is the difference of two counts that each carry a variance of about B, which is where the 2 under the root comes from:

LC = k√(2B) = 1.645 × √240 = 1.645 × 15.49 = 25.5 counts.

This is the decision threshold, applied after the count. A sample with no Cs-137 in it would exceed it in only 5% of counts.

Step 2: the detection limit LD

LD = k² + 2LC = 2.706 + 2 × 25.5 = 53.7 counts.

That is the true net signal that would clear LC in 95% of counts. It describes the method, so it is worked out from the blank and needs no sample result. At k = 1.645 the expression is the familiar 2.71 + 4.65√B.

Step 3: from counts to becquerels

Each becquerel gives ε·Y·t counts: Y photons per decay, ε counts per photon, t seconds. Here that is 0.030 × 0.851 × 3600 = 91.9 counts per becquerel, so

MDA = LD / (ε·Y·t) = 53.7 / 91.9 = 0.584 Bq.

For the 0.5 kg sample that is 1.17 Bq/kg. To reproduce this on the calculator, note that a fixed count has one efficiency field, “Detector efficiency”, described as counts per emission from the source, and no field for emission probability or sample mass. Enter the product of ε and Y, 2.553%, as that efficiency, with 2 cpm, 3600 s and the 95% / 95% option, and it returns the same LC, LD and 0.584 Bq. Divide by the mass yourself.

Reading a result

LC is what a result is compared with. The MDA is not a reporting threshold, and treating it as one is a frequent error. Three possible net counts from this measurement:

Net countsActivityConclusion
200.218 BqNot detected. Below L_C.
400.435 BqDetected, though below L_D.
700.762 BqDetected, and above L_D.

The middle row surprises people. A net count of 40 is a detection even though its 0.435 Bq is below the 0.584 Bq MDA, because the MDA describes what the method reliably catches, not what it is allowed to report. In the first row, keep the measured value and its uncertainty alongside the words “not detected” instead of writing zero. MARLAP asks laboratories to report each result and its uncertainty as obtained, even a negative one, and never to report a result as “less than MDC”. That also keeps later averages honest.

Paired blank or well-known blank

The 2 in √(2B) assumes a paired blank: one blank counted for the same time as the sample, carrying as much counting noise as the sample does. Count the blank for longer and its share of the noise shrinks. With B still scaled to the sample's count time, MARLAP gives the variance of a net count on an empty sample as B(1 + tS/tB), and then LC = k√(B(1 + tS/tB)) and LD = k² + 2LC. A blank known so well that it adds nothing is Currie's “well-known” case. His table gives 3.29σB for it, and with the k² term kept that is about 2.71 + 3.29√B. A blank shorter than the sample pushes the other way.

BlankLCLDMDA
Shorter, tB = tS/440.2983.30.906 Bq
Paired, tB = tS25.4853.70.584 Bq
Longer, tB = 4tS20.1543.00.468 Bq
Well-known18.0238.70.422 Bq

The MDA calculator implements only the paired row. If your blank is longer, its figure errs high and you can work the lower one from the expressions above. If your blank is shorter than the sample, its figure errs low, here by about 36%, and the unsafe direction is the one to watch. That row is only a guide, since its blank collects about 30 counts, below the range noted further down, where these expressions start to give too many false positives. Be careful about claiming the well-known row too. A background that drifts, or a blank from a different day, adds variation these Poisson expressions do not contain, and the lower figure is then not earned.