Lu-177 decay calculator: a worked example

Lu-177 decay calculation: a 7.4 GBq dose worked through

By Radiation Lab · Published · Reviewed

A Lu-177 therapy dose is prescribed for the moment of infusion, but it is calibrated days earlier and may sit in a vial for days afterwards. This note follows one dose of 7.4 GBq (200 mCi) from calibration to storage: how much activity has to be there at the start, what is in the vial on the day, what a delay costs, and what is left a week or two later. The percentages in the two tables below are entries of the Lu-177 remaining-activity table, and any other interval can be run in the decay calculator.

Which half-life, and what it means per day

This site uses 6.6443 days (159.46 hours), the evaluated value in the IAEA Livechart. European product information prints 6.647 days. Over 14 days the two differ by 0.06%, so the choice does not matter for planning, as long as one value is used throughout.

The shortcut to avoid is “a half-life is about a week”. After 7 days, 48.2% is left, not 50%. A more useful pair of figures for checking arithmetic in your head: Lu-177 loses 9.9% in the first day and 0.43% in the first hour.

How much activity is needed at calibration

Going back in time means dividing. If 7.4 GBq is wanted at infusion and t is the time from calibration to infusion, the calibrated activity is 7.4 / 2−t/T½:

Calibration to infusionRemainingNeeded at calibration
24 h90.1%8.21 GBq
48 h81.2%9.12 GBq
72 h73.1%10.12 GBq
96 h65.9%11.23 GBq

A ready-to-use product may already be labelled for the infusion time: the European information for lutetium oxodotreotide states 7 400 MBq at the date and time of infusion, from a fixed 370 MBq/mL at calibration with the volume set to suit. In that case the correction runs from the time on that label. The table is for a pharmacy working from a bulk calibration or checking a supplier's figure, and it is decay arithmetic only: the same information limits that product to 72 hours from calibration, so a row in this table is not a statement that a preparation can still be used that late.

Monday's calibration, Thursday's infusion

A vial is labelled 10.1 GBq at 10:00 on Monday. The infusion is booked for 10:00 on Thursday, with 7.4 GBq prescribed. What should the vial hold?

  1. Elapsed time: 3 days exactly.
  2. Half-lives: 3 / 6.6443 = 0.4515.
  3. Remaining fraction: 2−0.4515 = 0.7313, the 73.1% of the table's 3 d row.
  4. Activity: 10.1 × 0.7313 = 7.39 GBq, which is 99.8% of the prescription.

Open the calculator with these inputs to see the same result. Treat that figure as what the dose calibrator ought to show before administration. The calibrator reading is the measurement; the calculation tells you whether it looks right.

The same vial answers the opposite question. The “When does it reach” mode, with 10.1 GBq as the start and 7.4 GBq as the target, gives 2.982 days, or 71.56 hours, which is about 09:34 on Thursday. Before that moment the vial holds more than the prescription, and afterwards less.

Drawing from a bulk vial

Decay lowers the concentration while the volume stays where it was, so a dose drawn from bulk solution needs more millilitres as the days go by. Suppose the bulk is 2.0 GBq/mL at Monday's calibration. On Thursday the concentration is 2.0 × 0.7313 = 1.463 GBq/mL, and 7.4 GBq takes 5.06 mL. Drawing the 3.70 mL that Monday's concentration suggests would deliver 5.41 GBq, 26.9% short. A calculated volume is only a starting point; the syringe still goes in the calibrator.

What a delay costs

Take a vial holding exactly 7.4 GBq at the planned time. If the patient is two hours late it holds 7.34 GBq, 0.87% less. If treatment moves to the next day it holds 6.67 GBq, 9.9% less.

To see how far that goes, borrow a tolerance. The US label for lutetium Lu 177 dotatate supplies the vial at 7.4 GBq ± 10% at the time of injection, which makes the limits 6.66 and 8.14 GBq for this example. Solving t = T½·log₂(A₀/A) for each one, the activity falls through the upper limit 21.9 hours before the planned time and through the lower limit 24.2 hours after it, a span of about 46 hours, and the next-day case above only just stays inside.

That span describes activity and nothing more. It does not mean a dose can be moved by that much. A supply specification is not a dosing tolerance, and that same label limits its product to a shelf life of 72 hours from calibration. The Monday vial in this note is a generic arithmetic example rather than that product, but whatever you are using will have its own limit. Whether to delay or reschedule is decided by the product label, your protocol and the treating team, not by this arithmetic.

Clocks deserve a separate check. The European labelling for lutetium oxodotreotide prints these times in UTC. Read as local time, a UTC time is nine hours out at UTC+9: calibration really happened later than you assumed, so the vial holds 4.0% more than the calculation says. At UTC−5 the error runs the other way, five hours and 2.1% less. The larger of those would use 40% of the upper side of a ±10% band before anything else has gone wrong.

The vial after 7 and 14 days

A cancelled dose stays radioactive for a long time. These are the same table rows applied to 7.4 GBq:

Days after the 7.4 GBq timeRemainingActivity
0100.0%7.40 GBq
190.1%6.67 GBq
281.2%6.01 GBq
373.1%5.41 GBq
748.2%3.57 GBq
1423.2%1.72 GBq

So one week leaves 3.57 GBq and two weeks leave 1.72 GBq. These figures describe a vial and nothing else. Activity in a patient also falls through excretion, so it drops faster than physical decay alone, and this calculation does not model that.

When Lu-177m decides storage

Ten half-lives of Lu-177 take 66.4 days and reduce 7.4 GBq to 7.23 MBq. Whether that ends the matter depends on how the Lu-177 was made. Product information names two starting materials, lutetium-176 and ytterbium-176, with different waste management. Material made directly from Lu-176 can carry the isomer Lu-177m, half-life 160.4 days. Over those same 66.4 days the isomer only falls to 75.0%, while Lu-177 on its own has fallen below 0.1%. Waste that looks finished by the Lu-177 arithmetic may not be.

The decay calculator follows one nuclide at a time, so it will not show you this mix, nor the part of the isomer's decay that feeds back into Lu-177 by isomeric transition. Run Lu-177m as its own calculation, starting from the content your supplier states, read the supplier's documentation on waste, and apply the clearance rules that hold where you work. The Lu-177 sheet describes the two production routes.

Check it yourself

Sources